Appendix B: Status and Error Messages

        This appendix describes the status and error messages that appear on
        the CTS850 display. Status messages are statements about actions the
        CTS850 has completed. Status messages appear during normal
        operation. Error messages are statements about actions the CTS850
        was unable to complete. Error messages appear when the CTS850 is
        unable to perform a requested action.


Status Messages
        Autoscan complete. The CTS850 has successfully completed the
        Autoscan sequence.

        Burst pointer movement transmitted. The POINTER ACTION button
        was pressed with the Pointer Control set to Burst.

        Disk root directory read. The CTS850 has successfully read the disk
        directory.

        Factory settings restored, non-volatile settings cleared. This message
        appears when the instrument is turned on after it was turned off
        during a state change. The instrument settings were corrupted during
        the shutdown and so the factory default settings were loaded at
        turn-on and all information stored in RAM was cleared.

        File has been deleted. The CTS850 has successfully deleted the
        selected file.

        Hardcopy complete. The CTS850 has successfully completed a
        printout.




CTS850 SDH Test Set User Manual                                             B 1
Appendix B: Status and Error Messages




         Instrument factory reset complete. The recall of default factory
         settings, from the RECALL INSTRUMENT SETUPS page, was
         completed successfully.

         Instrument returned to previous setup. This message appears if you
         select Cancel while an Autoscan is in progress.

         Instrument setup for received signal. The CTS850 has completed an
         Autoscan and has changed its settings to match those of the received
         signal.

         Internal diagnostics completed. An instrument self test has been
         initiated and completed successfully.

         No Setup file with '.SET' extension found. You selected the Disk action
         on the RECALL INSTRUMENT SETUPS page and no instrument
         setup files were found on the disk.

         Previous power down instrument state restored. This message appears
         when the CTS850 turns on. It means that the settings in effect when
         the CTS850 was last turned off have been successfully restored.

         Reading disk.... The CTS850 is reading the disk directory.

         Recall of instrument setup complete. This message confirms that the
         recall of an instrument setup was successfully completed.

         Recall of measurement results/histories complete. The CTS850 has
         successfully recalled the selected results from disk.

         Recall of pass/fail test complete. The CTS850 has successfully
         recalled the selected pass/fail test from disk.

         Save of instrument setup complete. This message confirms that an
         instrument setup was successfully written to disk.




B 2                                         CTS850 SDH Test Set User Manual
                                       Appendix B: Status and Error Messages




        Save of measurement results/histories complete. The CTS850 has
        successfully saved test results to disk.

        Save of pass/fail test complete. The CTS850 has successfully saved
        the pass/fail test to disk.

        Single pointer movement transmitted. You pressed the POINTER
        ACTION button with the Pointer Control set to Single.

        Test restarted due to change in receiver settings. You changed the
        receive settings while a test was running. When any settings are
        changed, a running test is restarted.

        Transmit rate change complete. The CTS850 has successfully changed
        transmit rate. This message can occur when an instrument setup file
        is recalled from disk.


Error Messages
        Alarm generation invalid while in through mode. You attempted to
        transmit an alarm while the CTS850 was in Through Mode. Transmit
        settings cannot be changed while the transmit rate is set to Through
        Mode.

        Autoscan already in progress. You pressed the Autoscan front-panel
        button while an Autoscan was in progress.

        Button disabled in edit mode. You pressed a front-panel button while
        the CTS850 was in edit mode. You must exit edit mode before you
        can use the front-panel buttons.

        Button disabled when dialog box is displayed. You pressed a
        front-panel button while a dialog box (for example, PRINT
        CONTROL) was displayed. Some front-panel buttons are disabled
        while a dialog box is displayed. Exit the dialog box to enable all the
        front-panel buttons.



CTS850 SDH Test Set User Manual                                              B 3
Appendix B: Status and Error Messages




         Can not insert tributary error/alarm while adding external signal. An
         attempt was made to insert a PDH error or alarm while adding an
         external signal.

         Could not create disk file. The CTS850 was unable to create a file on
         the disk. Try saving the file on a different disk.

         Could not delete file. You attempted to a delete a file and a disk error
         occurred that prevented the CTS850 from deleting the file. The file is
         probably corrupted, though you might be able to fix the problem with
         an MS-DOS disk recovery program.

         Could not open disk file. The CTS850 has attempted to read a file
         from disk and was unable to do so. The file is probably corrupted,
         though you might be able to recover the file using an MS-DOS file
         recovery program.

         Could not read disk's root directory. The CTS850 was unable to read
         the disk directory. The disk directory is probably corrupted, though
         you might be able to recover files from the disk with an MS-DOS
         disk recovery utility.

         Diagnostics invalid while disk or autoscan busy. You tried to run Self
         Test diagnostics while accessing the disk or completing an Autoscan.
         Wait until disk access is complete or the Autoscan is complete before
         running a self test routine.

         Disk file too large for memory buffer, truncated. You recalled a file and
         it was too large to load into memory. This can occur if an instrument
         settings file is recalled that has been edited on a PC and the edited
         file is larger than the original file.

         Disk has changed, operation aborted. You attempted to recall a disk
         file and the disk has been changed since the directory was read.
         Reinsert the original disk and recall the file again.

         Disk is busy. You tried to save a file to disk while the disk was active.
         Wait until the disk activity light turns off and save to disk again.


B 4                                          CTS850 SDH Test Set User Manual
                                         Appendix B: Status and Error Messages




        Disk is full. You tried to save a file to disk and there is not enough
        room on the disk for the file. Insert a disk with space available and
        save the file again.

        Disk not present in drive. You tried to save a file to disk, but no disk is
        in the drive. Insert a formatted disk and save the file again.

        Disk is write-protected, cannot write to file. You tried to save a file to a
        disk that is write-protected.

        Disk operation failed. A disk error occurred while reading or writing
        to the disk. If you were trying to save a file to disk, save the file to a
        different disk. If you were trying to read a file from disk, you might
        be able to recover the file using an MS-DOS disk recovery program.

        Error insertion disabled while in Through Mode. You pressed the
        INSERT ERROR button while the CTS850 was in Through Mode.
        You cannot insert errors while the CTS850 is in Through Mode
        because the transmit signal must match the received signal.

        Errors in test recall, test not loaded. The CTS850 encountered disk
        errors while recalling a pass/fail test and the operation was aborted.
        You might be able to recover the file with an MS-DOS disk recovery
        program.

        Extra data in history file ignored. You recalled a history file that
        contains extra data. This can occur if the history file has been edited
        with a file editor and the edited file contains inappropriate informa-
        tion.

        Failure generation invalid while in through mode. You attempted to
        transmit a failure while the CTS850 was in Through Mode. Transmit
        settings cannot be changed while the transmit rate is set to Through
        Mode.

        File name required. You entered a name for an instrument setup, a
        pass/fail test, or a results file and then you removed the name by
        selecting Clear. This message is just a reminder that a name must be


CTS850 SDH Test Set User Manual                                                  B 5
Appendix B: Status and Error Messages




         entered for the file to be saved to disk; that is, there is no default file
         name.

         Filename not found. This message appears if the disk was swapped
         after the directory was read and you attempted to recall a file. Insert
         the original disk and recall the file again.

         File not read. A disk error occurred while reading a file. The file is
         probably corrupted, though you might be able to recover the file with
         an MS-DOS disk recovery program.

         Frequency offset disabled with current transmit clock. You attempted
         to change from Pointer/Timing Mode to Frequency Offset while the
         transmit clock was set to External. Frequency Offset can be selected
         only when the transmit clock is set to Internal or Recovered.

         Hardcopy already in progress. You tried to start a printout while a
         printout was already in progress. Wait until the current printout
         completes and then select Print again.

         Instrument unable to drop signal while transmitting current rate. You
         attempted to turn on tributary drop while the transmit rate was set to
         a PDH rate. Because the PDH transmit connector and tributary drop
         connector are the same connector, you cannot transmit a PDH signal
         and drop a PDH signal at the same time.

         Internal diagnostics failed. A malfunction has occurred that requires
         servicing. Contact a Tektronix representative for assistance.

         Internal Error N: Contact Tektronix Service. A hardware malfunction
         has occurred that requires servicing. Contact a Tektronix representa-
         tive for assistance.

         Internal format of file incorrect; file not read. You recalled a file that
         does not contain data in the correct format. This occurs when a file is
         recalled that happens to have the same extension as the type of
         CTS850 file being recalled, but it is not a CTS850 file.



B 6                                           CTS850 SDH Test Set User Manual
                                       Appendix B: Status and Error Messages




        Invalid history data format in file. You recalled a file that does not
        contain data in the correct format. This occurs when a file is recalled
        that happens to have the correct extension for a history file (and thus
        shows up on the RECALL RESULTS page) even though it is not a
        CTS850 history file.

        Loss of added signal. You have selected an externally generated
        tributary signal, but no valid signal is present.

        Loss of external clock. You have selected an external clock source
        that does not have a valid clock source attached.

        Measurements are running, recall not allowed. You tried to recall
        instrument setups from memory or disk while a test is running. Wait
        until the test is completed and then recall the file.

        Measurements are running, results cannot be saved. You attempted to
        save results to disk while a test was running. Wait until the test is
        completed and then save the results to disk.

        No Result file with '.RES' extension found. This message appears when
        you select Disk (on the RECALL RESULTS page) but the disk does
        not contain any results files.

        Not available without tributary option. The requested action requires
        that the Add/Drop/Test tributary option be installed.

        No Test file with '.TST' extension found. This message appears when
        you select Disk (on the RECALL PASS/FAIL TESTS page) but the
        disk does not contain any pass/fail test files.

        Numeric value greater than maximum limit. You attempted to enter a
        numeric value larger than the maximum value.

        Numeric value less than minimum limit. You attempted to enter a
        numeric value smaller than the minimum value.




CTS850 SDH Test Set User Manual                                                 B 7
Appendix B: Status and Error Messages




         Operation invalid while a test is running. You tried to perform an
         action that cannot be completed while a test is running. For example,
         this message would appear if you attempted to recall instrument
         settings while a test is running.

         Optical power has been overloaded. You have connected an optical
         signal that has saturated the CTS850 optical receiver.

         Pointer burst active, request ignored. You pressed the POINTER
         ACTION button while a pointer burst movement was occurring.

         Pointer movements disabled while in Through Mode. You pressed the
         POINTER ACTION button while the CTS850 was in Through Mode.
         Pointer movements are not allowed while the CTS850 is in Through
         Mode because the transmit signal must match the received signal.

         Root directory is full. You tried to save a file to disk but the directory
         shows that the disk is full. Insert a disk with space available and save
         the file again.

         Through Mode invalid while receiving a tributary signal. You attempted
         to change the transmit rate to Through Mode while a test was
         running at a tributary rate. Any change in settings would invalidate
         the test, so no change was made.

         Unexpected End Of File, file not read. You tried to recall an instrument
         setup file, a pass/fail file, or a history file and a disk error occurred.
         The file is corrupted and cannot be recovered.

         Warning: No received signal currently identified. You started a test
         while there was no received signal.

         Write Failure on Disk. The CTS850 was unable to write to the disk.
         Try saving the file on a different disk.




B 8                                           CTS850 SDH Test Set User Manual