Service Manual




SD-22 & SD-26
Sampling Head
070-7227-02




Warning
The servicing instructions are for use by qualified
personnel only. To avoid personal injury, do not
perform any servicing unless you are qualified to
do so. Refer to the Safety Summary prior to per-
forming service.

Please check for change information at the rear
of this manual.

First Printing: May 1993

www.tektronix.com
Instrument Serial Numbers
Each instrument manufactured by Tektronix has a serial number on a panel insert or tag, or stamped on the
chassis. The first letter in the serial number designates the country of manufacture. The last five digits of the
serial number are assigned sequentially and are unique to each instrument. Those manufactured in the
United States have six unique digits. The country of manufacture is identified as follows:
    B010000          Tektronix, Inc., Beaverton, Oregon, USA
    E200000          Tektronix United Kingdom, Ltd., London
    J300000          Sony/Tektronix, Japan
    H700000          Tektronix Holland, NV, Heerenveen, The Netherlands

Instruments manufactured for Tektronix by external vendors outside the United States are assigned a two digit
alpha code to identify the country of manufacture (e.g., JP for Japan, HK for Hong Kong, IL for Israel, etc.).
Tektronix, Inc., P Box 500, Beaverton, OR 97077
                  .O.
Printed in U.S.A.
Copyright E Tektronix, Inc., 1989, 1993. All rights reserved. Tektronix products are covered by U.S. and
foreign patents, issued and pending. The following are registered trademarks: TEKTRONIX, TEK,
TEKPROBE, and SCOPE MOBILE.
                                                        WARRANTY
Tektronix warrants that this product will be free from defects in materials and workmanship for a period of one (1) year from the
date of shipment. If any such product proves defective during this warranty period, Tektronix, at its option, either will repair the
defective product without charge for parts and labor, or will provide a replacement in exchange for the defective product.

In order to obtain service under this warranty, Customer must notify Tektronix of the defect before the expiration of the warranty
period and make suitable arrangements for the performance of service. Customer shall be responsible for packaging and
shipping the defective product to the service center designated by Tektronix, with shipping charges prepaid. Tektronix shall
pay for the return of the product to Customer if the shipment is to a location within the country in which the Tektronix service
center is located. Customer shall be responsible for paying all shipping charges, duties, taxes, and any other charges for
products returned to any other locations.

This warranty shall not apply to any defect, failure or damage caused by improper use or improper or inadequate maintenance
and care. Tektronix shall not be obligated to furnish service under this warranty a) to repair damage resulting from attempts by
personnel other than Tektronix representatives to install, repair or service the product; b) to repair damage resulting from
improper use or connection to incompatible equipment; or c) to service a product that has been modified or integrated with
other products when the effect of such modification or integration increases the time or difficulty of servicing the product.

THIS WARRANTY IS GIVEN BY TEKTRONIX WITH RESPECT TO THIS PRODUCT IN LIEU OF ANY OTHER
WARRANTIES, EXPRESSED OR IMPLIED. TEKTRONIX AND ITS VENDORS DISCLAIM ANY IMPLIED WARRANTIES OF
MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. TEKTRONIX' RESPONSIBILITY TO REPAIR OR
REPLACE DEFECTIVE PRODUCTS IS THE SOLE AND EXCLUSIVE REMEDY PROVIDED TO THE CUSTOMER FOR
BREACH OF THIS WARRANTY. TEKTRONIX AND ITS VENDORS WILL NOT BE LIABLE FOR ANY INDIRECT, SPECIAL,
INCIDENTAL, OR CONSEQUENTIAL DAMAGES IRRESPECTIVE OF WHETHER TEKTRONIX OR THE VENDOR HAS
ADVANCE NOTICE OF THE POSSIBILITY OF SUCH DAMAGES.
                                                                                    Table of Contents

                               List of Figures . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .      iii
                               List of Tables . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .      v




General Information
                               General Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .             11
                                     Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .      11
                                     Safety Summary . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .            12
                                     Installing and Removing the Sampling Head . . . . . . . . . . . . . . .                                     14
                                     Packaging for Shipment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .                  16
                                     Operating Environment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .                 17
                                         Operating Temperature . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .                   17




Checks and Adjustments
                               Checks and Adjustments . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .                  21
                                     Test Equipment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .          23
                                     Using These Procedures . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .                    25
                                         Conventions in this Manual . . . . . . . . . . . . . . . . . . . . . . . . . . . . .                    25
                                         Initialized and Stored Settings . . . . . . . . . . . . . . . . . . . . . . . . . .                     25
                                         Menu Selections and Measurement Techniques . . . . . . . . . . .                                        25
                                         Setup Illustrations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .           26
                                     Part 1 Power On . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .           28
                                     Part 2 Dot Transient Response . . . . . . . . . . . . . . . . . . . . . . . . . . . .                      2 10
                                     Part 3 Offset . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .    2 16
                                     Part 4 Noise . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .     2 20
                                     Part 5 Rise Time . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .         2 23
                                     Part 6 Acquisition Aberrations . . . . . . . . . . . . . . . . . . . . . . . . . . . .                     2 26
                                     Part 7 Coincidence Between Channels . . . . . . . . . . . . . . . . . . . .                                2 30
                                     Part 8 Maximum Signal Voltage . . . . . . . . . . . . . . . . . . . . . . . . . . .                        2 33
                                     Part 9 Isolation Between Channels . . . . . . . . . . . . . . . . . . . . . . . .                          2 36




Maintenance
                               Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .       31
                                     Preventive Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .                  31
                                        Periodic Electrical Checks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .                   31



SD 22 & SD 26 Service Manual                                                                                                                       i
Table of Contents



                                  Static Sensitive Device Classification . . . . . . . . . . . . . . . . . . . .                         31
                                  Exchanging Sampling Heads . . . . . . . . . . . . . . . . . . . . . . . . . . .                        33
                                  Removing and Replacing the Sampling Head Internal
                                      Circuitry . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .      34
                                  Changing the Sampling Head Identification Number . . . . . . .                                         36




Theory of Operation
                      Theory of Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .              41
                            System Functional Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .                       41
                            Loop Gain . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .      42
                            Offset Null . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .    43




Replaceable Parts
                      Replaceable Parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .            51
                            Parts Ordering Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .                     51
                            Using the Replaceable Parts List . . . . . . . . . . . . . . . . . . . . . . . . . .                         52




ii                                                                                                                                  Contents
                                                                                                 List of Figures

                               Figure 1 1: Sampling Head Compartments in an 11801B and a
                                   CSA 803A . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .        14
                               Figure 1 2: Installing a Sampling Head in an Instrument . . . . . . . . . .                                       15
                               Figure 2 1: Locations of Controls and Connectors on Mainframe
                                   Instruments . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .       2   7
                               Figure 3 1: Sampling Head Screw Locations . . . . . . . . . . . . . . . . . . . .                                 3   5
                               Figure 3 2: A5 Time Base/Controller Board Jumper Location . . . . . .                                             3   7
                               Figure 4 1: SD 22 and SD 26 Sampling Head Block Diagram . . . . . .                                               4   1
                               Figure 4 2: Displayed Traces at Various Loop Gain Settings . . . . . . .                                          4   2
                               Figure 5 1: Exploded View of the SD 22 and SD 26 Sampling
                                   Heads . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .   55




SD 22 & SD 26 Service Manual                                                                                                                     iii
List of Figures




iv                Contents
                                                                                               List of Tables

                               Table 2   1:   Measurement Limits and Specifications . . . . . . . . . . . . . . .                       21
                               Table 2   2:   Test Equipment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .    23
                               Table 2   3:   Aberration Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . .        2 26
                               Table 3   1:   Relative Susceptibility to Electrostatic Discharge (ESD) . .                              32




SD 22 & SD 26 Service Manual                                                                                                             v
List of Tables




vi               Contents
General Information
                                                         General Information

                               This section gives all the information needed to apply power to the SD 22 &
                               SD 26 Sampling Heads.
                               Safety information as well as information on installing and removing the
                               sampling head, packaging for shipment, and environmental conditions such
                               as operating temperature is included here.




Introduction                   The SD 22 & SD 26 Sampling Head Service Manual is designed for use by
                               qualified service personnel. It contains information necessary to check and
                               maintain the SD 22 & SD 26 Sampling Heads.
                               The SD 22 Sampling Head is a two channel, low noise, 28 ps rise time
                               sampling head. The SD 26 Sampling Head is a two channel, 17.5 ps rise
                               time sampling head. The SD 22 & SD 26 Sampling Heads are designed for
                               use in the 11800 Series Digital Sampling Oscilloscopes, the SM 11 Multi
                               Channel Unit, and the CSA 803 Series Communications Signal Analyzers.




SD 22 & SD 26 Service Manual                                                                             11
General Information




                      The general safety information in this part of the summary is for both operat
Safety Summary
                      ing and servicing personnel. Specific warnings and cautions will be found
                      throughout the manual where they apply, but may not appear in this
                      summary.


                      Terms in Manuals
                      CAUTION statements in manuals identify conditions or practices that could
                      result in damage to the equipment or other property.
                      WARNING statements in manuals identify conditions or practices that could
                      result in personal injury or loss of life.


                      Terms on Equipment
                      CAUTION on equipment means a personal injury hazard not immediately
                      accessible as one reads the marking, or a hazard to property including the
                      equipment itself.
                      DANGER on equipment means a personal injury hazard immediately acces
                      sible as one reads the marking.


                      Symbols in Manuals



                                              Static Sensitive Devices


                      Symbols on Equipment



                              DANGER                Protective           ATTENTION
                             High Voltage         ground (earth)           Refer to
                                                     terminal              manual


                      Grounding the Instrument
                      The sampling head is grounded through the grounding conductor of the
                      oscilloscope's power cord. To avoid electric shock, plug the power cord into
                      a properly wired receptacle before connecting to the product input or output
                      terminal. A protective ground connection, by way of the grounding conduc
                      tor in the power cord, is essential for safe operation.


                      Danger Arising from Loss of Ground
                      Upon loss of the protective ground connection, all accessible conductive
                      parts (including knobs and controls that may appear to be insulating), can
                      render an electric shock.



12                                                                               General Information
                                                                                      General Information



                               Do Not Operate in Explosive Atmospheres
                               To avoid explosion, do not operate the sampling head in an atmosphere of
                               explosive gasses.




SD 22 & SD 26 Service Manual                                                                          13
General Information




Installing and        To avoid damage to the instrument, set the instrument's ON/STANDBY
                      switch to STANDBY before installing or removing the sampling head.
Removing the
Sampling Head         The SD 22 & SD 26 Sampling Head slide into the one of the front panel
                      compartments of the 11800 Series Digital Sampling Oscilloscopes or CSA
                      803 Series Communications Signal Analyzers. Figure 1 1 shows the front
                      panel of an instrument and the locations of the sampling head
                      compartments.


                                                            Sampling Head
                                                            Compartments

                                     11801B




                                                                    Sampling Head
                                                                    Compartments

                                     CSA 803A




                                                      Power Only Sampling
                                                      Head Compartments


                           Figure 1 1: Sampling Head Compartments in an 11801B and a
                                                   CSA 803A




14                                                                           General Information
                                                                                           General Information



                                With the ON/STANDBY switch set to STANDBY, place the sampling head in a
                                compartment and slowly push it in with firm pressure. Once the sampling
                                head is seated, turn the lock down screw to tighten the sampling head into
                                place. See Figure 1 2.


                                                                  CAUTION


                                     To prevent damage to your instrument or sampling head, never
                                     install or remove a sampling head when the ON/STANDBY switch
                                     is ON.


                                     If the green indicator light remains on when the STANDBY position
                                     is selected, then the switch has been left internally disabled after
                                     the servicing of the power supply. To enable the ON/STANDBY
                                     switch, refer to the Maintenance section of the Service Manual for
                                     your mainframe instrument.


                                To remove the sampling head from an instrument, set the instrument's
                                ON/STANDBY switch to STANDBY. Turn the lock down screw and then
                                slowly pull out the sampling head.




                                                                            Mainframe




                               Lock Down
                                 Screw




                                                       Sampling
                                                         Head


                                           Figure 1 2: Installing a Sampling Head in an Instrument




SD 22 & SD 26 Service Manual                                                                                15
General Information




Packaging for         If possible, save and reuse the original carton and packaging to the pack
                      age sampling head when shipping it by commercial transportation.
Shipment
                      Attach a tag to the sampling head if it is shipped to your local Tektronix
                      service center for service or repair. Include the following information on the
                      tag:
                      H   Name and address of the instrument and sampling head owner.
                      H   Name of a person at your firm who can be contacted about the instru
                          ment and sampling head.
                      H   Complete instrument and sampling head type and serial number.
                      H   A description of the service required.
                      Package the sampling head as follows, if the original package is not avail
                      able or is not fit for use:

                          Step 1: Attach short circuit terminations to the sampling head inputs.

                          Step 2: Obtain a corrugated cardboard carton with inside dimensions at
                          least six inches (15 cm) greater than the sampling head dimensions.
                          Use a carton with a bursting test strength of at least 200 pounds per
                          square inch.

                          Step 3: Fully wrap the sampling head with antistatic sheeting, or its
                          equivalent, to protect the finish.

                          Step 4: Tightly pack dunnage or urethane foam between the carton and
                          the sampling head to cushion the sampling head on all sides. Allow
                          three inches of packing on each side.

                          Step 5: Seal the carton with shipping tape or with industrial staples.

                          Step 6: Mark the address of the Tektronix service center and your return
                          address on the carton in one or more prominent places.




16                                                                                General Information
                                                                                        General Information




Operating                      The following environmental requirements are provided to ensure proper
                               operation and long sampling head life.
Environment
                               Operating Temperature
                               Operate the sampling head where the ambient air temperature is between
                               0_ C and +50_ C. Store the sampling head in ambient temperatures from
                               -40_ C to +75_ C. After storage at temperatures outside the operating
                               limits, allow the chassis to reach the operating temperature range before
                               applying power.
                               Allow a 20 minute warmup before performing the calibrations in the
                               Enhanced Accuracy menu of the instrument mainframe. Calibrations should
                               be repeated if the temperature changes more than "5