Service Manuals, User Guides, Schematic Diagrams or docs for : Tektronix TEK P7240 USER

<< Back | Home

Most service manuals and schematics are PDF files, so You will need Adobre Acrobat Reader to view : Acrobat Download Some of the files are DjVu format. Readers and resources available here : DjVu Resources
For the compressed files, most common are zip and rar. Please, extract files with Your favorite compression software ( WinZip, WinRAR ... ) before viewing. If a document has multiple parts, You should download all, before extracting.
Good luck. Repair on Your own risk. Make sure You know what You are doing.




Image preview - the first page of the document
TEK P7240 USER


>> Download TEK P7240 USER documenatation <<

Text preview - extract from the document
User Manual




P7240
4 GHz 5X Active Probe
071-0759-03




Warning
The servicing instructions are for use by qualified
personnel only. To avoid personal injury, do not
perform any servicing unless you are qualified to
do so. Refer to all safety summaries prior to
performing service.

www.tektronix.com
CopyrightE Tektronix, Inc. All rights reserved.
Tektronix products are covered by U.S. and foreign patents, issued and
pending. Information in this publication supercedes that in all previously
published material. Specifications and price change priviledges reserved.
Tektronix, Inc., P Box 500, Beaverton, OR 97077
                  .O.
TEKTRONIX, TEK, and SureFoot are registered trademarks, and TekConnect,
SureToe, and KlipChip are trademarks of Tektronix, Inc.
                                 WARRANTY
Tektronix warrants that this product will be free from defects in materials and
workmanship for a period of one (1) year from the date of shipment. If any such
product proves defective during this warranty period, Tektronix, at its option,
either will repair the defective product without charge for parts and labor, or will
provide a replacement in exchange for the defective product.
In order to obtain service under this warranty, Customer must notify Tektronix of
the defect before the expiration of the warranty period and make suitable
arrangements for the performance of service. Customer shall be responsible for
packaging and shipping the defective product to the service center designated
by Tektronix, with shipping charges prepaid. Tektronix shall pay for the return of
the product to Customer if the shipment is to a location within the country in
which the Tektronix service center is located. Customer shall be responsible for
paying all shipping charges, duties, taxes, and any other charges for products
returned to any other locations.
This warranty shall not apply to any defect, failure or damage caused by im-
proper use or improper or inadequate maintenance and care. Tektronix shall not
be obligated to furnish service under this warranty a) to repair damage resulting
from attempts by personnel other than Tektronix representatives to install, repair
or service the product; b) to repair damage resulting from improper use or
connection to incompatible equipment; or c) to service a product that has been
modified or integrated with other products when the effect of such modification
or integration increases the time or difficulty of servicing the product.
THIS WARRANTY IS GIVEN BY TEKTRONIX WITH RESPECT TO THIS
PRODUCT IN LIEU OF ANY OTHER WARRANTIES, EXPRESSED OR IM-
PLIED. TEKTRONIX AND ITS VENDORS DISCLAIM ANY IMPLIED WARRAN-
TIES OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE.
TEKTRONIX' RESPONSIBILITY TO REPAIR OR REPLACE DEFECTIVE
PRODUCTS IS THE SOLE AND EXCLUSIVE REMEDY PROVIDED TO THE
CUSTOMER FOR BREACH OF THIS WARRANTY. TEKTRONIX AND ITS
VENDORS WILL NOT BE LIABLE FOR ANY INDIRECT, SPECIAL, INCIDEN-
TAL, OR CONSEQUENTIAL DAMAGES IRRESPECTIVE OF WHETHER
TEKTRONIX OR THE VENDOR HAS ADVANCE NOTICE OF THE POSSIBIL-
ITY OF SUCH DAMAGES.
General Safety Summary
        Review the following safety precautions to avoid injury and prevent
        damage to this product or any products connected to it. To avoid
        potential hazards, use this product only as specified.
        Only qualified personnel should perform service procedures.

        To Avoid Fire or Personal Injury
        Connect and Disconnect Properly. Connect the probe output to the
        measurement instrument before connecting the probe to the circuit
        under test. Disconnect the probe input and the probe ground from the
        circuit under test before disconnecting the probe from the measure-
        ment instrument.
        Observe All Terminal Ratings. To avoid fire or shock hazard, observe
        all ratings and markings on the product. Consult the product manual
        for further ratings information before making connections to the
        product.
        Connect the ground lead of the probe to earth ground only.
        Do Not Operate Without Covers. Do not operate this product with
        covers or panels removed.
        Do Not Operate With Suspected Failures. If you suspect there is
        damage to this product, have it inspected by qualified service
        personnel.
        Do Not Operate in Wet/Damp Conditions.
        Do Not Operate in an Explosive Atmosphere.
        Keep Product Surfaces Clean and Dry.




P7240 4 GHz 5X Active Probe User Manual                                        i
General Safety Summary




        Safety Terms and Symbols
        Terms in This Manual. These terms may appear in this manual:

        WARNING. Warning statements identify conditions or practices that
        could result in injury or loss of life.


        CAUTION. Caution statements identify conditions or practices that
        could result in damage to this product or other property.


        Terms on the Product. These terms may appear on the product:
        DANGER indicates an injury hazard immediately accessible as you
        read the marking.
        WARNING indicates an injury hazard not immediately accessible as
        you read the marking.
        CAUTION indicates a hazard to property including the product.
        Symbols on the Product. These symbols may appear on the product:



            CAUTION
          Refer to Manual




ii                              P7240 4 GHz 5X Active Probe User Manual
Table of Contents

        General Safety Summary . . . . . . . . . . . . . . . . . . . . . . . . . . . .                    iii
        Preface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .    v
        Related Manuals . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .            v
        Contacting Tektronix . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .            vi
        Product Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .              1
        TekConnect Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .             1
        Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .     2
        Options . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .    3
        Features and Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . .                 5
        Functional Check . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .            13
        Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .         15
        Probe Offset . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .      15
        Operating Basics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .            17
        Handling the Probe . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .            17
        Maximum Nondestructive Input Voltage . . . . . . . . . . . . . . . . .                            17
        Input Linear Dynamic Range . . . . . . . . . . . . . . . . . . . . . . . . . .                    17
        Ground Lead Length . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .              18
        Ground Lead Inductance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .                18
        Helpful Hints . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .         21
        Low-inductance Grounding . . . . . . . . . . . . . . . . . . . . . . . . . . . .                  21
        SureFoot Grounding . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .              22
        Appendix A: Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . .                  23
        Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .      23
        Appendix B: User Service . . . . . . . . . . . . . . . . . . . . . . . . . . . .                  27
        Inspection and Cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .               27
        Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .     27
        Replacement Parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .           27
        Preparation for Shipment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .              28
        Appendix C: Replaceable Parts . . . . . . . . . . . . . . . . . . . . . . .                       29
        Parts Ordering Information . . . . . . . . . . . . . . . . . . . . . . . . . . . .                29
        Using the Replaceable Parts List . . . . . . . . . . . . . . . . . . . . . . . .                  30




P7240 4 GHz 5X Active Probe User Manual                                                                   iii
Table of Contents




List of Figures

         Figure 1: P7240 Probe featuring the TekConnect interface                          1
         Figure 2: Connecting and disconnecting the probe . . . . . . .                     2
         Figure 3: Probe functional check connections . . . . . . . . . . .                13
         Figure 4: Dynamic and offset limitations . . . . . . . . . . . . . . .            16
         Figure 5: Waveform distortion from ground lead length . .                         18
         Figure 6: Ground Lead Equivalent Circuit . . . . . . . . . . . . . .              19
         Figure 7: Low-inductance grounding . . . . . . . . . . . . . . . . . . .          21
         Figure 8: Using a SureFoot adapter for grounding . . . . . . .                    22
         Figure 9: Typical input impedance and
            phase versus frequency . . . . . . . . . . . . . . . . . . . . . . . . . . .   24
         Figure 10: Typical bandwidth . . . . . . . . . . . . . . . . . . . . . . . . .    25
         Figure 11: Probe head and compensation box dimensions .                           25
         Figure 12: P7240 replaceable parts . . . . . . . . . . . . . . . . . . . .        31
         Figure 13: P7240 standard accessories . . . . . . . . . . . . . . . . .           32
         Figure 14: P7240 optional accessories . . . . . . . . . . . . . . . . . .         34




iv                                      P7240 4 GHz 5X Active Probe User Manual
Preface

        This is the User Manual for the P7240 probe. This manual provides
        operating information, specifications, and a replaceable parts list.


Related Manuals
        If you need to do a performance verification or make internal
        adjustments to your probe, refer to the P7240 4 GHz Active Probe
        Service Manual, Tektronix part number 071-1056-XX. The manual is
        a printable pdf file, and is available on both the Tektronix website
        and the Optional Applications CD, Tektronix part number
        063-3376-XX. The Optional Applications CD is included with
        Tektronix oscilloscopes featuring the TekConnect interface.




P7240 4 GHz 5X Active Probe User Manual                                        v
Preface




Contacting Tektronix


          Phone         1-800-833-9200*


          Address       Tektronix, Inc.
                        Department or name (if known)
                        14200 SW Karl Braun Drive
                        P.O. Box 500
                        Beaverton, OR 97077
                        USA


          Web site      www.tektronix.com


          Sales         1-800-833-9200, select option 1*
          support


          Service       1-800-833-9200, select option 2*
          support


          Technical     Email: [email protected]
          support
                        1-800-833-9200, select option 3*
                        1-503-627-2400
                        6:00 a.m. - 5:00 p.m. Pacific time

          *   This phone number is toll free in North America. After office
              hours, please leave a voice mail message.
              Outside North America, contact a Tektronix sales office or
              distributor; see the Tektronix web site for a list of offices.




vi                               P7240 4 GHz 5X Active Probe User Manual
Product Description

        The Tektronix P7240 is a 4 GHz (probe only), 5X active FET probe.
        The P7240 has a low input capacitance (1 pF) and high input
        resistance (20 k), which minimize circuit loading over a wide
        bandwidth range. The small profile and low-mass head of the P7240
        make probing dense circuitry by hand fast and easy. The accessory
        tips and adapters included enable the P7240 to be used on a wide
        variety of circuit architectures.




        Figure 1: P7240 Probe featuring the TekConnect interface


TekConnect Interface
        The P7240 is powered through a TekConnect interface between the
        probe compensation box and the oscilloscope. The TekConnect
        interface provides a communication path through contact pins on the
        host instrument. Power, signal, offset, and probe characteristic data
        transfer through the interface.




P7240 4 GHz 5X Active Probe User Manual                                     1
Product Description




         When the probe is connected, the oscilloscope reads EEPROM
         information from the probe, identifying the device and allowing the
         appropriate power supplies to be powered on. The preamp inputs on
         the oscilloscope are electrostatic discharge protected by remaining
         grounded until a valid TekConnect device is detected.


Installation
         The TekConnect interface features a spring-loaded latch that
         provides audible and tactile confirmation that a reliable connection
         has been made to the oscilloscope. Slide the probe into the
         TekConnect receptacle on the oscilloscope. The probe snaps into the
         oscilloscope when fully engaged. See Figure 2.
         To release the probe from the oscilloscope, grasp the compensation
         box, depress the latch button, and pull out the probe.


                                         Latch
                                         button




         Figure 2: Connecting and disconnecting the probe




2                                P7240 4 GHz 5X Active Probe User Manual
                                                       Product Description




Options
        The following options are available when ordering the P7240 probe:
        H Option D1 - Calibration Data
        H Option C3 - 3 years Calibration Service
        H Option D3 - 3 years Calibration Data (requires Option C3)
        H Option R3 - 3 years Extended Warranty




P7240 4 GHz 5X Active Probe User Manual                                      3
Product Description




4                     P7240 4 GHz 5X Active Probe User Manual
Features and Accessories

             The P7240 is provided with several features and accessories designed
             to make probing and measurement a simpler task. To familiarize
             yourself with these items and their uses, refer to Table 1.

Table 1: P7240 features and standard accessories

Feature/Accessory                         Description
                                          Probe head assembly. The probe head is
                                          designed for ease of use and high performance.
                                          Its small size makes it easy to handle in tight
                                          areas.
 Probe tip
  socket                                  The probe tip socket is sized to easily press onto
                          Ground          0.025 inch square pins for direct access. The
                          socket          ground socket provides a short ground path for
                                          high fidelity ground connections.
     Stabilization
        notch                             The stabilization notch permits you to use
                                          adjacent pins to reduce stresses on the probe
                                          and pins.
                                        TekConnect interface. The TekConnect
                                 Latch interface provides a communication path between
                                 button the probe and the oscilloscope. Contact pins
                                        provide power, signal, offset, and probe
                                        characteristic data transfer.
                                          The probe snaps into the oscilloscope when fully
                                          engaged. To remove, grasp the compensation
                                          box, depress the latch button, and pull the probe
                                          out.
                                          Push-in probe tip (10 ea). Use the push-in
                                          probe tip for general purpose probing by hand.
                                          The push-in probe tip may also be used with the
                                          other socketed leads and adapters.
                      Push-in probe tip
                                          Tektronix part number: 131-5638-11
                                          (package of 10)


P7240 4 GHz 5X Active Probe User Manual                                                        5
Features and Accessories




Table 1: P7240 features and standard accessories (Cont.)

Feature/Accessory                     Description
                                      Installing the push-in probe tip. Attach the
                                      push-in probe tip by aligning the tip into the probe
                                      tip socket and pushing the tip in until it is seated.
                                      Either end of the tip may be used.
                                      Do not force the tip. Also, be careful not to poke
                                      yourself with the sharp probe tip. To remove the
                                      tip, gently grab the tip with small pliers, and pull
                                      the tip out.
                                      SureToe probe tip (4 ea). The SureToe tip is a
                                      pointed probe tip useful for probing in dense
                                      circuitry. Attach the SureToe adapters the same
                                      way as the push-in probe tips.
                                      Do not force the tip. Also, be careful not to poke
                                      yourself with the sharp probe tip. They can be
                                      used with any of the socketed accessory leads.
                                      Tektronix part number: ST501 (package of 12)
                                      SMT KlipChip (2 ea). Use the SMT KlipChip
                                      test clips to access fragile, dense circuitry.
        KlipChip
                                      KlipChip test clips can be connected to the Y-lead
                                      or three-inch ground leads. Simply press the lead
                                      socket into the KlipChip handle.
                                      The KlipChip body freely turns, allowing better
                    Y-lead adapter    probe orientation. To reduce stress and provide a
                                      lower profile on components being tested, the
                                      flexible sleeve of the KlipChip bends up to a
                                      35 degree angle.
                                      Tektronix part number: 206-0364-XX




6                                    P7240 4 GHz 5X Active Probe User Manual
                                                         Features and Accessories




Table 1: P7240 features and standard accessories (Cont.)

Feature/Accessory                   Description
                                    Y-lead adapter (2 ea). Use the Y-lead adapter
                   Y-lead           to extend the physical reach of the probe and
                  adapter
                                    ground when necessary. The Y-lead adapter
                                    accepts any of the probe tips or adapters, and
                                    can be pushed directly onto 0.025 inch square-
                                    pins.
                                    When selecting the grounding connection,
                                    maintain as short a ground path as possible.
                                    Refer to page 18 for more information.
                                    To attach the Y-lead adapter, gently press the
         Y-lead                     lead pins into the probe head tip and ground
        adapter                     receptacles. Using the black lead for ground is
                                    recommended.




                                    Three-inch ground lead (3 ea). Use the
                                    three-inch ground lead for general probing. The
                                    socketed end of the lead may be connected to
                                    any of the probe tips and adapters or fitted onto
                                    0.025 inch square pins.
                                    To attach the ground lead, press and rotate the
                                    lead pin connector into the ground socket on the
                                    probe head. The lead may be removed by simply
                                    pulling the pin out by hand.
                                    When selecting the grounding connection,
                                    maintain as short a ground path as possible.
                                    Refer to page 18 for more information.
                                    Tektronix part number: 196-3456-00 (package of
                                    three ground leads and two Y-lead adapters)




P7240 4 GHz 5X Active Probe User Manual                                                 7
Features and Accessories




Table 1: P7240 features and standard accessories (Cont.)

Feature/Accessory                   Description
                                    Customizable ground lead (5 ea). This
                                    ground lead wire can be bent or cut shorter.
                                    NOTE: To ease insertion into the ground socket
                                    of the probe, cut the tip of this ground lead wire at
                                    a 30 to 60 degree angle.
                                    To maintain signal fidelity while probing, use as
              Customizable          short a ground path as possible. Refer to page 18
              ground lead           for more grounding information.
                                    Tektronix part number: 196-3482-00
                                    (package of 5)
                                    Adapter spring (10 ea). Use the adapter spring
                                    for low-profile probing of 0.025 inch square pins.
                                    The adapter spring allows the probe to lie at a
                                    right-angle (flat against a circuit board). This
                                    enables probing in vertical circuits, such as
                                    computer or communications backplanes, or in
                                    tight areas, such as between circuit cards.
                                    The adapter can be used directly with the probe
                                    head or attached to the Y-lead adapter or ground
                                    leads.

                        Right       The adapter is attached by pushing the tip into
                        angle       the probe tip socket until it is seated. The adapter
                        adapter     spring can be easily removed by hand.
                                    Tektronix part number: 016-1774-00
                                    (package of 10)




8                                  P7240 4 GHz 5X Active Probe User Manual
                                                            Features and Accessories




Table 1: P7240 features and standard accessories (Cont.)

Feature/Accessory                     Description
                                      Low-inductance ground pogo pin (10 ea).
                                      Use the low-inductance ground pogo pin to
                                      substantially reduce ground lead inductance.
                                      Because the pogo pin simply touches the ground
                                      reference, you can easily move the probe to
                                      different points on the circuit under test.
                                      To attach, press the pogo pin into the probe head
  Pogo pin                            ground socket.
                                      When selecting the grounding connection,
                                      maintain as short a ground path as possible.
                                      Refer to page 18 for more information.
                                      Tektronix part number: 016-1772-10
                                      (package of 10)
                                      Ground pin length adapter (10 ea).
                                      The ground pin length adapter is ideal for use
                                      with signal/ground pairs on 0.100 inch square-
                                      header pins.
                                      Attach the adapter by gently pressing it into the
                                      ground socket on the probe head.
                       Ground pin     Be sure to use the stabilization notch whenever
                       length adapter possible to avoid slipping and damaging the
                                      probe or circuitry under test.
                                      Tektronix part number: 016-1773-10
                                      (package of 10)
                     Cable marker     Cable marker bands (10 ea). Attach matching
                     bands            pairs of the marker bands onto the cable at the
                                      head and compensation box of each probe. The
                                      marker bands enable quick verification of which
                                      probe is connected to which instrument channel.

                                      Tektronix part number: 016-1315-00
                                      (package of 10)




P7240 4 GHz 5X Active Probe User Manual                                                   9
Features and Accessories




Table 1: P7240 features and standard accessories (Cont.)

Feature/Accessory                   Description
                                    Antistatic wrist strap. When using the probe,
                                    always work at an antistatic work station and
                                    wear the antistatic wrist strap.



                                    Tektronix part number: 006-3415-04
                                    Plastic accessory box. Use the plastic box to
                                    store the probe accessories when not in use.

                                    Tektronix part number: 006-7164-00
                                    Instrument case. The instrument case protects
                                    the probe from harsh environments and provides
                                    room for storing optional accessories.

                                    Tektronix part number: 016-1879-01
                                    Instruction Manual. Provides specifications and
                                    instructions for operating the probe, and a list of
                                    accessories and adapters.

                                    Tektronix part number: 071-0759-XX
                                    Calibration certificate. A certificate of traceable
                                    calibration is provided with every instrument
                                    shipped.




10                                 P7240 4 GHz 5X Active Probe User Manual
                                                           Features and Accessories




         Table 2 lists the optional accessories you can order for your P7240
         probe.

Table 2: P7240 optional accessories

Accessory                             Description
                                      SureFoot probe tip. The SureFoot tip is an
                                      integral probe tip and miniature guide that
                                      enables fault-free probing of fine-pitch SMD
                                      packages.
                                      Attach SureFoot tips the same way as the push-in
                                      probe tips. Three models of SureFoot tips are
                                      included with the probe and can be used with any
                                      of the socketed accessory leads.
                                      SF501. The orange SureFoot tip is compatible
                                      with 50 mil JEDEC packages such as SOIC,
                                      PLCC, CLCC, and so on.
                                      Tektronix part number: SF501
                                      (package of 12)
                                      SF502. The blue SureFoot tip is compatible with
                                      0.65 mm JEDEC and EIAJ packages.
                                      Tektronix part number: SF502
                                      (package of 12)
                                      SF503. The red SureFoot tip is compatible with
                                      EIAJ packages.
                                      Tektronix part number: SF503
                                      (package of 12)
                                      Micro KlipChip adapters (4 ea). Use the
                                      adapters to probe the leads on integrated circuits
                                      that are surface-mounted.
                                      Tektronix part number: SMK4
                                      (package of 4)




P7240 4 GHz 5X Active Probe User Manual                                                 11
Features and Accessories




Table 2: P7240 optional accessories (Cont.)

Accessory                            Description
                                     SMA-to-probe tip adapter. Use the adapter to
                                     connect the probe to SMA cables. The adapter
                                     includes a 50-ohm SMA terminator.




                                     Tektronix part number: 015-0678-00
                                     IEEE1394 Adapter. The IEEE1394 Adapter
                                     allows you to probe signals on the bus, external
                                     to system enclosures, without disturbing system
                                     operation. The adapter maintains a balanced
                                     55-ohm signal path and can be used in both
                                     single-ended and differential modes.
                                     Tektronix part number: 679-5027-00
                                     TekConnect interface calibration fixture. The
                                     calibration fixture is required when a performance
                                     verification or adjustment is done on the probe. It
                                     provides connectors and test points for internal
                                     probe measurements.

                                     Tektronix part number: 067-0422-00
                                     Calibration software and instructions.
                                     Use the Optional Applications Software CD that is
                                     shipped with oscilloscopes featuring the
                                     TekConnect interface. Alternatively, you can
                                     download the software from the Tektronix
                                     website, or order the CD using the part number
                                     below.
                                     Tektronix part number: 063-3376-XX




12                                 P7240 4 GHz 5X Active Probe User Manual
Functional Check

        A functional check may be performed using the PROBE COM-
        PENSATION connections on the front panel of the oscilloscope. See
        Figure 3.




        Figure 3: Probe functional check connections


        1. Connect the probe to a TekConnect receptacle on the oscillo-
           scope.
        2. Set the oscilloscope to display the appropriate probe channel.
        3. Using a ground lead and SMT KlipChip, connect the probe
           ground to the outer ring of the PROBE COMPENSATION
           connector on the oscilloscope.
        4. Using a standard tip, hold the probe to the center conductor of the
           PROBE COMPENSATION BNC connector on the oscilloscope.
        5. Press AUTOSET, or adjust the oscilloscope to display a stable
           calibration waveform. (You may need to adjust the probe offset to
           display the waveform.)


P7240 4 GHz 5X Active Probe User Manual                                     13
Functional Check




        NOTE. Now is a good time to perform a probe calibration routine
        from your instrument.

        6. Disconnect the probe tip from the oscilloscope, and ground the
           probe tip. (Connect the KlipChip to the probe tip.)
        7. With the probe offset set to 0.0 V, the oscilloscope display should
           be at the ground reference.
        8. Set the oscilloscope volts/division to 500 mV.
        9. Adjust the probe offset. The displayed waveform should vary
           between approximately +1.0 V and - 1.0 V.




14                               P7240 4 GHz 5X Active Probe User Manual
Configuration

        The P7240 provides the oscilloscope with the probe model number,
        serial number, and attenuation factor. When connected to an
        oscilloscope with a TekConnect interface, display readouts are
        corrected for the probe attenuation factor, the instrument input is set
        to 50 , and the coupling is set to DC. The probe offset control is
        controlled by the oscilloscope.


Probe Offset
        The probe offset is adjustable to permit operation within the linear
        range of the probe. Using the offset to cancel DC signal components
        enables optimal probe performance. See Figure 4 on page 16 for
        more information.


        NOTE. See your oscilloscope manual for specific instructions on its
        operation and offset control.

        To set the probe offset, follow these steps:
        1. Ground the input of the probe.
        2. Use the vertical position control to set a zero reference level on
           the oscilloscope display.
        3. Set the oscilloscope to 1 V/div.
        4. Attach the probe to the circuit.
        5. Adjust the probe offset to bring the trace to the oscilloscope zero
           reference.
        6. Change the volts/division setting to the desired range, adjusting
           the offset to keep the trace on the zero reference level.




P7240 4 GHz 5X Active Probe User Manual                                         15
Configuration




             NOTE. The P7240 has a 