            Reference




            TDS5000 Series
            Digital Phosphor Oscilloscopes
            071-1020-00




*P071102000*
071102000
To Use the Front Panel
You can use the dedicated, front-panel knobs and buttons to do the most common operations.

   Turn INTENSITY to adjust waveform intensity.

           Push FastAcq to turn Fast Acquisition
                                     on and off.
      Push CURSORS to turn cursors on and off.

               Push PRINT to make a hard copy.

  Push DEFAULT SETUP to return settings to the
                      factory default values.

 Push AUTOSET to automatically set the vertical,
    horizontal, and trigger controls to produce a
                                  usable display.



          Use the multipurpose knobs to control
  parameters selected from the screen interface.
 Push a FINE button to make small changes with
                          a multipurpose knob.



             Push OFF to turn the optional touch
                              screen on and off.

  Push ZOOM to add a magnified graticule to the
     display. Push the HORIZ or VERT button to
        select the axis that you want to magnify.


  Use these knobs and buttons to set horizontal
 SCALE and POSITION for the waveforms. Push
DELAY to turn on horizontal delay, and then use
        POSITION to set the delay time. Adjust
RESOLUTION to change the number of acquired
                          points in the waveform.


    Use these knobs and buttons to set the basic
 trigger parameters. Push ADVANCED to display
           a menu of additional trigger functions.


            Use these buttons to start and stop
          acquisition or start a single acquisition
      sequence. The ARM, READY, and TRIG'D
              lights show the acquisition status.



Push CH to turn channel displays on and off.
    Adjust the channel POSITION and SCALE or
    change the input termination with dedicated
                            knobs and buttons.




Copyright E Tektronix, Inc. Beaverton, OR 97077
To Use the Screen Interface
You can control all oscilloscope functions, including powering off the instrument, through the screen interface.

                   Choose the Toolbar
    Click here to change to menu-bar operation.



        Click a button in the toolbar to display a
     control window at the bottom of the display.




      Click a screen control to change a setting.



  Click a numerical control to assign that control
  to a multipurpose knob. Turn the multipurpose
             knob to adjust the parameter value.




           Or Choose the Menu Bar
      Use some menu items to display a control
     window at the bottom or side of the display.


                Use some menu items to directly
                             change settings.



       Click here to change to toolbar operation.




            Click here to close a control window.



                                                            More Operating Tips:
                                                            H Use the optional touch screen to control the oscilloscope when bench space is
                                                              unavailable, such as on a cart or in an equipment rack.

                                                            H Plug in a mouse and keyboard if you have the bench space to use them. You can
                                                              plug in a USB mouse or keyboard anytime, even while the oscilloscope is
                                                              running.

                                                            H Use the menu bar to access PC-related functions such as, Page Setup, Export,
                                                              and Copy.
To Display a Waveform

 Attach a probe to CH 1 and connect the probe        1
                               to your signal.




                   Push CH 1 if channel 1 is not     2
                            already displayed.

                                Push AUTOSET.        3




           Adjust VERTICAL and HORIZONTAL            4
          POSITION and SCALE if necessary to
                            optimize the display.



    Adjust RESOLUTION to change the record           5
     length and sample rate. You can acquire
   more samples in the waveform to see more
       detail, or acquire fewer samples with a
                             faster update rate.

   Adjust INTENSITY to change the brightness,        6
 vector fill, and display persistence of acquired
                                           points.

              Recommended Probe
                    Accessories
                                                         P5050 Passive Probes for                P6243 and P6245 Active Probes or P6158
                                                         general-purpose applications            Low-C Probe for high-speed applications




                                                         P6246, P6247, P6248 Differential Probes TCP202 Current Probe for
                                                         for differential signals and low-noise  general-purpose applications
                                                         applications
To See More Waveform Detail

                                Use Zoom               Use the Zoom function to magnify an acquisition vertically, horizontally, or in both
                                                       waveform dimensions. SCALE or POSITION changes you make to the Zoom graticule
                                                       affect only the display, not the actual waveform that is acquired.

             Push the ZOOM button to split the     1
             screen and add a zoom graticule.

  Push the HORIZ button or the VERT button to      2
select the axis you want to magnify in the zoom                                                     Main graticule       Zoom graticule
graticule. Use the multipurpose knobs to adjust
 scale and position of the magnified waveform.


          You can also set up a zoom graticule     3
      from the screen interface. Click and drag
      across the segment of the waveform that
              you want to see in greater detail.

   Then select Zoom from the drop-down list to     4
   magnify the highlighted waveform segment.




               Use Horizontal Delay                    Use horizontal DELAY to acquire waveform detail in a region that is separated from
                                                       the trigger location by a significant interval of time.
                                                            Trigger point
                                                                                                          Acquired waveform




                                                                                    Delay time


                                                                                                            Expansion point
           Push the front-panel DELAY button.      1
       Adjust the delay time with the horizontal   2
      POSITION control, or enter the delay time
                         in the control window.


        Adjust the horizontal SCALE to acquire     3
         the detail you need around the delay
                               expansion point.


                                                       More Operating Tips:
                                                       H You can use Zoom and Horizontal Delay together to magnify a delayed
                                                         acquisition.

                                                       H Toggle Horizontal Delay on and off to quickly compare signal details at two
                                                         different areas of interest, one near the trigger location and the other centered at
                                                         the delay time.
To Use Fast Acquisition
Turn Fast Acquisition on to acquire up to 100,000 waveforms per second.



             Push the front-panel FastAcq button.       1
                 Or select Fast Acquisitions in the
                                 Horiz/Acq menu.




    Adjust INTENSITY to optimize the intensity or       2
color grading for the signal you want to analyze.
   Increasing intensity can make less-frequently
          acquired points brighter in the display.




                         How DPOs Work
Unlike DSOs that have long dead times between
    acquisitions, Digital Phosphor Oscilloscopes
(DPOs) continuously acquire waveforms at rates
     comparable to analog oscilloscopes. DPOs
   continuously overlay the acquired information
        into a three-dimensional database that is
updated on the display 30 times per second. For
  each pixel in the display, the intensity (or color)
      of the pixel is proportional to the number of
         actual samples that the pixel represents.




                        Analog Real Time sweeps



                       Digital Storage acquisitions


                     Digital Phosphor acquisitions




                                                            More Operating Tips:
                                                            H Choose one of the color grading palettes in the Display Colors control window to
                                                              see different sample densities represented in different colors.

                                                            H Turn AutoBright on in the Display Appearance control window. When you use
                                                              AutoBright, the displayed waveforms remain visible even at low trigger repetition
                                                              rates.
To Choose an Acquisition Mode




               Select Acquisition Mode... in the   1
                               Horiz/Acq menu.

      Or click the Horiz button, and then select
                     an acquisition mode in the
         horizontal/acquisition control window.




  For Average or Envelope acquisition modes,       2
    click the # of Wfms control and then set the
   number of waveforms with the multipurpose
knob. You can also double-click the control and
                        use the pop-up keypad.


                                                                                     record duration                  Displayed record points (at
                 How the Acquisition                    acquisition interval +
                                                                                 number of points in record        maximum horizontal magnification)
                        Modes Work
                                                       Interval 1    2            3         4                     Interval 1   2        3        4

  Sample mode retains one sampled point from
                      each acquisition interval.
                                                                                                      Sample



       Peak Detect mode uses the highest and
     lowest of all the samples contained in two
              consecutive acquisition intervals.                                                   Peak Detect




      Hi Res mode calculates the average of all
      the samples for each acquisition interval.                                                      Hi Res


                                                                    Three acquisitions from one source


                                                         Acquisition 1                 2                      3
Envelope mode finds highest and lowest record
 points over many acquisitions. Envelope uses
   Peak Detect for each individual acquisition.                                                                                    Envelope


Average mode calculates the average value for
     each record point over many acquisitions.
                                                                                                                                   Average
Average uses Sample mode for each individual
                                  acquisition.
To Select a Trigger




  Select the EDGE trigger type and then set the
   source, coupling, slope and mode with these
front-panel controls. Push ADVANCED to select
                  one of the other trigger types.


       You can also select a trigger type in the
                                 Trigger menu.


      Or click the Trig button and then select a
      trigger type in the trigger control window
                                that is displayed.
Trigger Selections

 Trigger type   Levels                     Timers                      Trigger conditions

 Edge           Single level               None                        Trigger on rising or falling edge, as defined by slope
                                                                       control. Coupling choices are DC, AC, AC LF Reject, AC
                                                                       HF Reject, and Noise Reject.

 Video          Single level               None                        Trigger on NTSC, PAL, SECAM, Analog HDTV, and
                                                                       Custom video formats.


 Glitch         Single level               One to specify glitch       Trigger on glitches narrower than the specified width or
                                           width                       ignore glitches narrower than the specified width.


 Width          Single level               Two to specify minimum      Trigger on pulses that have widths between the range of
                                           and maximum pulse           the two timers or outside the range of the two timers.
                                           widths


 Runt           Two levels to define the   One to specify an           Trigger on a pulse that enters the transition region from
                logic transition region    optional minimum            one side but does not leave the region from the other side.
                                           runt-pulse duration         Pulses can be qualified with the logical states of other
                                                                       channels (4-channel models only).

 Window         Two levels to define       One to specify an           Trigger when a signal violates either upper or lower
                signal transition region   optional minimum            boundaries set outside the normal range of the signal.
                                           violation width             Events can be qualified with the logical states of other
                                                                       channels (4-channel models only).

 Timeout        Single level               One to specify time-out     Trigger when a signal does not make a transition for a
                                           time                        specified length of time.


 Transition     Two levels to define the   One to specify transition   Trigger when a logic signal spends more time or less time
                logic transition region    time                        in the transition region than a specified amount of time.


 Setup/Hold     Independent levels for     One to specify setup        Trigger on violations of setup or hold time between a Data
                Data and Clock             time and one to specify     signal and a Clock signal. The specified setup and hold
                                           hold time                   times can be positive or negative values.

 Pattern        Independent levels for     One to specify pattern      Trigger when a Boolean combination of up to four
                each channel               duration                    channels becomes true. Trigger immediately or only after
                                                                       the combination is true for a specified time duration.

 State          Independent levels for     None                        Trigger on transition of one channel when a Boolean
                each channel                                           combination of up to three other channels is true.
To Use A (Main) and B (Delayed) Triggers
You can use the A Event (Main) trigger alone or combine it with the B Event (Delayed) trigger to capture more complex signals.


         Set the A trigger type and source in the     1
         A Event (Main) tab of the trigger control
                                         window.




 Choose a function in the AB Sequence tab of          2
                      the trigger control window.


        Set the trigger delay time or the number      3
                     of B events, as appropriate.



  Set the B trigger characteristics in the B Event    4
     (Delayed) tab of the trigger control window.

                                                                                                                 Pretrigger                     Posttrigger
                                                                                                                 record                         record

                                                                                                                 Acquired waveform record
                    Trigger on B Event                                A trigger point

                                                          A trigger
            The A trigger arms the oscilloscope.           source
    Posttrigger acquisition starts on nth B event.
                                                          B trigger
                                                           source
                                                                                                                              B trigger point
                                                                                               Waiting for nth
                                                                                             event (Where n=5)


        B Trigger After Delay Time                                    A trigger point

            The A trigger arms the oscilloscope.          A trigger
      Posttrigger acquisition starts on the first B        source
               edge after the trigger delay time.
                                                          B trigger
                                                           source                               Trigger
                                                                                               delay time                     B trigger point




                                                           More Operating Tips:
                                                           H B-trigger delay time and horizontal delay time are independent functions. When
                                                             you establish a trigger condition using either the A trigger alone or the A and B
                                                             triggers together, you can also use horizontal delay to delay the acquisition by an
                                                             additional amount.

                                                           H When using the B trigger, the A trigger can be any of the following types: Edge,
                                                             Glitch, or Width. The B trigger type is always Edge type.
To Take Measurements with Cursors

       Push the front-panel CURSORS button.       1


Select the waveform you want to measure and       2
    a cursor type in the cursor control window.

   Or you can activate cursors on the selected
        waveform directly in the Cursor menu.




          Place cursors with the multipurpose     3
           knobs or enter the cursor locations
                                  numerically.




            If you choose split cursors to take   4
          measurements between waveforms,
            select the source for each cursor.                                                                 T1: 356.4 ms
                                                                                                               T2: 352.5 ms
                                                                                                               DT: 3.92 ms
                                                                                                               1/DT: 255 Hz
                                                                                                               V1: 5.120 V
                                                                                                               V2: 4.886 V
                                                                                                               DV: 234 mV
                                                                                                               DV/DT: 59.7 V/s




                   Read cursor measurement        5
                        results in the display.
                                                      Other Cursor Measurement Tips:
                                                      H You can set the cursors to move together in tandem if you choose the Tracking
                                                        cursor mode. The cursors move independently if you choose the Independent
                                                        cursor mode.

                                                      H If you use the zoom graticule, you can place a cursor directly on a specific
                                                        waveform point to take precision measurements.

                                                      H You can also move cursors by dragging them to a new position.
To Take Automated Measurements

   Click the Meas button, and then select up to
  eight measurements using the measurement
                               control window.




     Use the tabs to access measurements in
                      four different categories.




   Or choose a measurement for the selected
      waveform directly in the Measure menu.




Automated Measurement Selections
              Amplitude                                      Time                    More                 Histogram (Option 2A)



                               Positive                                              Burst      Wfm                            Std
  High           Mean         Overshoot        Period     Rise Time    Delay                                     Max
                                                                                     Width      Count                        Deviation


                              Negative                                 Positive                 Hits in
   Low          Pk