Text preview - extract from the document
                     TM 9-6625-474-14&P-2


            TECHNICAL   MANUAL


          OPERATOR, ORGANIZATIONAL,
     DIRECT SUPPORT, AND GENERAL SUPPORT
              MAINTENANCE MANUAL
            (INCLUDING REPAIR PARTS)
                      FOR



       FUNCTION GENERATOR
     TEKTRONIX, MODEL FG 501A
          (6625-01-106-9873)




DEPARTMENT OF THE ARMY       27 DECEMBER 1984
                                                                               TM 9-6625-474-14&P-2


                                            WARNING




                                                                         RA PD 404264


                                  DANGEROUS VOLTAGE
                          is used in the operation of this equipment
                                    DEATH ON CONTACT
                  may result if personnel fail to observe safety precautions



Never work on electronic equipment unless there is another person nearby who is familiar with the
operation and hazards of the equipment and who is competent in administering first aid. When the
technician is aided by operators, he must warn them about dangerous areas.
Whenever possible, the power supply to the equipment must be shut off before beginning work on the
equipment. Take particular care to ground every capacitor likely to hold a dangerous potential. When
working inside the equipment, after the power has been turned off, always ground every part before
touching it.
Be careful not to contact high-voltage connections when installing or operating this equipment.
Whenever the nature of the operation permits, keep one hand away from the equipment to reduce the
hazard of current flowing through vital organs of the body.


                                            WARNING

Do not be misled by the term "low voltage." Potentials as low as 50 volts may cause death under adverse
conditions.
COMMON and probe ground straps are electrically connected. Herefore, an elevated reference applied to
any is present on each - as indicated by the yellow warning bands under the probe retractable hook tips.
For Artificial Respiration, refer to FM 21-11,

                                            Power Source

This product is intended to operate in a power module connected to a power source that will not apply more
than 250 volts rms between the supply conductors or between either supply conductor and ground. A
protective ground connection by way of the grounding conductor in the power cord is essential for safe
operation.


                                                                                            a/(b blank)
                                                                                                             TM 9-6625-474-14&P-2


                                                  Copyright 1980 Tektronix, Inc. All rights reserved
                                                   Reproduced by permission of copyright owner.

TECHNICAL MANUAL                                                                                            HEADQUARTERS
                                                                                                       DEPARTMENT OF THE ARMY
No. 9-6625-474-14&P-2                                                                    Washington, D.C., 27 December 1984


                                  OPERATOR, ORGANIZATIONAL, DIRECT SUPPORT,
                                  AND GENERAL SUPPORT MAINTENANCE MANUAL
                                          (INCLUDING REPAIR PARTS)
                                                     FOR
                                                 FUNCTION GENERATOR
                                              TEKTRONIX, MODEL FG 501A
                                                  (6625-01-106-9873)

                                                          REPORTING OF ERRORS

                     You can help improve this manual. If you find any mistakes or if you know of a way
                     to improve the procedures, please let us know. Mail your letter or DA Form 2028
                     (Recommended Changes to Publications and Blank Forms), direct to: Command-
                     er, US Army Missile Command, ATTN: DRSMI-SNPM, Redstone Arsenal, AL
                     35898-5238. A reply will be furnished to you.


                                                            TABLE OF CONTENTS
                                                                                                                                     Page
LIST OF ILLUSTRATIONS                                                                                                                 iii
LIST OF TABLES                                                                                                                         iv

SECTION 0.                                                                                                                            0-1
                                                                                                                                      0-1
                                                                                                                                      0-1
                                                                                                                                      0-1
                                                                                                                                      0-1
                                                                                                                                      0-1
                                                                                                                                      0-1

SECTION 1.                                                                                                                            1-1

SECTION 2.                                                                                                                            2-1
                                                                                                                                      2-1
                                                                                                                                      2-2
                                                                                                                                      2-4
                                                                                                                                      2-5
                                                                                                                                      2-6
                                                                                                                                      2-9

 SECTION 3.                                                                                                                           3-1

 SECTION 4.                                                                                                                           4-1
                                                                                                                                      4-1
                                                                                                                                      4-8

 SECTION 5.                                                                                                                           5-1
                                                                                                                                      5-1
                                                                                                                                      5-4




 This manual is, in part, authenticated manufacturer's commercial literature. Recommended Spare Parts List has been added to supplement the commercial
 literature, The format of this manual has not been structured to consider levels of maintenance.




                                                                                                                                                     i
TM 9-6625-474-14&P-2

                                                       TABLE OF CONTENTS (CONT)
                                                                                                                                                                                                                                                                   Page


SECTION 6.    OPTIONS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .                                                                                                              6-1

SECTION 7.    REPLACEABLE              ELECTRICAL          PARTS,              .           .           .           .       .       .               .           .               .           .               .           .               .               .       .   7-1

SECTION 8.    DIAGRAMS            AND          CIRCUIT             BOARD                       ILLUSTRATIONS                                       .               .                   .                   .               .                       .           .   8-1

SECTION 9.    REPLACEABLE               MECHANICAL                 PARTS                   .               .           .           .                   .               .               .                   .                   .                   .           .   9-1

APPENDIX A.   REFERENCES           .   .   .   .   .   .   .   .      .    .       .       .       .       .       .   .       .       .       .           .   .           .       .           .       .           .       .           .       .           .   .   A-1

APPENDIX B.   MAINTENANCE                  ALLOCATION                     CHART                            (MAC)                           .                   .                       .                           .                       .                   .   B-1

APPENDIX C.   RECOMMENDED              SPARE       PARTS           LIST.....           .       .       .       .       .       .       .       .           .       .           .       .           .           .       .           .           .           .   .   C-1

APPENDIX D.   MANUAL CHANGE INFORMATION. . . . . . . . . . . . . . . . . . . . . .. . . . . . . . . . . . . . . . . . . . . . . . . .. . . . . . . . . . .                                                                                                         D-1




ii
                                                                                                                   TM 9-6625-474-14&P-2


                                                     LIST OF ILLUSTRATIONS

Fig. No.                                                               Title

0-1        FG 501A 2 MHz Function Generator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .                          0-2
2-1        Installation and Removal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .              2-1
2-2        Controls and Connectors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .             2-3
2-3        Swept Frequency Range (VCF) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .                   2-6
2-4        Basic Functions (Waveforms) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .                 2-7
2-5        Ramps and Pulses . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .          2-7
2-6        Phase Relationships (Output and Trig) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .                         2-7
2-7        Trigger Signal Amplitude Requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .                           2-7
2-8        Gated Operation (Waveforms) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .                   2-8
2-9        Triggered Operation (Waveforms) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .                     2-8
2-10       Phase Control (Waveforms) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .                 2-8
2-11       Analyzing Circuit or System Response . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .                          2-9
2-12       Tone Burst Generation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .           2-10
4-1        Test Setup for Dial Alignment and Offset Adjustment . . . . . . . . . . . . . . . . . . . . .                                       4-9
4-2        Test Setup for Sine Distortion Adjustment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .                           4-11
4-3        Test Setup for Offset and Sine/SquareAmplitude
              Adjustment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .   4-12
4-4        Test Setup for Squarewave Comp/Rise and Falltime
              Adjustment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .   4-13
4-5        Test Setup for Dial Cal and Loop Delay Adjustment . . . . . . . . . . . . . . . . . . . . . . .                                     4-14
5-1        Coaxial End Lead Connector Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .                           5-3
5-2        Orientation and Disassembly of Multipin Connectors . . . . . . . . . . . . . . . . . . . . . .                                      5-3
5-3        Extension Shaft and Pushbutton Removal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .                              5-4
5-4        Rear Interface Connector Assignments . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .                          5-5




                                                                                                                                                      iii
TM 9-6625-474-14&P-2


                                                                  LIST OF TABLES

     Table No.                                                               Title                                                                 Page

       1-1       Electrical Characteristics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .        1-2
       1-2       Miscellaneous . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .   1-4
       1-3       Environmental . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .   1-4
       1-4       Physical Characteristics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .        1-5
       4-1       Test Equipment Required . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .             4-1
       5-1       Relative Susceptibility to Static Discharge . . . . . . . . . . . . . . . . . . . . . . . . . . . . .                             5-1




iv
                                                                                TM 9-6625-474-14&P-2


                                                  SECTION 0

                                           GENERAL INFORMATION


0-1. Scope. This manual contains instructions for        0-4. Reporting Equipment Improvement Recom-
the operator, organizational, direct support, and gen-   mendations (EIR). If your FG 501A needs im-
eral support maintenance of and calibration proce-       provement, let us know. Send us an EIR. You, the
dures for Tektronix Function Generator, Model FG         user, are the only one who can tell us what you don't
501A. Throughout this manual, Tektronix Function         like about your equipment. Let us know why you don't
Generator, Model FG 501A is referred to as the FG        like the design. Tell us why a procedure is hard to
501A.                                                    perform. Put it on an SF 368 (Quality Deficiency
                                                         Report). Mail it to Commander, U.S. Army Missile
0-2. Indexes of publications. a. DA Pam 310-4.           Command, ATTN: DRSM1-CIMD, Redstone Arsenal,
Refer to the latest issue of DA Pam 310-4 to deter-      AL 35898-5290. We'll send you a reply.
mine whether there are new editions, changes, or
additional publications pertaining to Tektronix Func-    0-5. Administrative Storage. To prepare the Tek-
tion Generator, Model FG 501A.                           tronix Function Generator, Model FG 501A for place-
   b. DA Pam 310-7. Refer to the latest issue of DA      ment into and removal from administrative storage,
Pam 310-7 to determine whether there are modifica-       refer to Section 3, Chapter 4, AR 750-25-1, Mainte-
tion work orders (MWO'S) pertaining to Tektronix         nance of Equipment and Supplies. Temporary storage
Function Generator, Model FG 501A.                       should be accomplished in accordance with TB 750-
                                                         25-1, Section 2, Maintenance of Supplies and
0-3. Forms, Records, and Reports. Department of          Equipment.
Army forms and procedures used for equipment main-
tenance and calibration are those prescribed by TM       0 - 6 . D e s t r u c t i o n o f A r m y Electronics
38-750, The Army Maintenance Management System.          Materiel. Destruction of Tektronix Function
Accidents involving injury to personnel or damage to     Generator, Model FG 501A to prevent enemy use shall
materiel will be reported on DA Form 285, Accident       be in accordance with TM 750-244-2.
Report, in accordance with AR 385-40.




                                                                                                          0-1
TM 9-6625-474-14&P-2




                                                                  .




                       FIG.0-1. FG 501A 2MHz FUNCTION GENERATOR




0-2
                                                                                          TM 9-6625-474-14&P-2

                                                        SECTION 1

                                      SPECIFICATION
                 INTRODUCTION                                                    ACCESSORIES
    This section of the manual contains a general descrip-        The only accessory shipped with the FG 501A is the
tion of the FG 501A and complete electrical, environmen-       Instruction Manual.
tal, and physical specifications. Standard accessories are
also listed. Instrument option information is located in the
back of this manual in a separate section.                            PERFORMANCE CONDITIONS
                                                                  The electrical characteristics are valid with the follow-
                                                               ing conditions:
        INSTRUMENT DESCRIPTION
                                                                 1. The instrument must have been adjusted at an
    The FG 501A Function Generator provides low distor-        ambient temperature between +20