>> Download sc504_cal documenatation <<Text preview - extract from the document *TB 9-6625-2173-35
DEPARTMENT OF THE ARMY TECHNICAL BULLETIN
CALIBRATION PROCEDURE FOR
OSCILLOSCOPE
TEKTRONIX, TYPES SC501, SC502, AND SC504
Headquarters, Department of the Army, Washington, DC
7 January 2004
Distribution Statement A: Approved for public release; distribution is unlimited.
REPORTING OF ERRORS AND RECOMMENDING IMPROVEMENTS
You can improve this manual. If you find any mistakes or if you know of a way to
improve these procedures, please let us know. Mail your letter or DA Form 2028
(Recommended Changes to Publications and Blank Forms) directly to: Commander,
US Army Aviation and Missile Command, AMSAM-MMC-MA-NP, Redstone Arsenal,
AL 35898-5000. A reply will be furnished to you. You may also provide DA Form
2028 information to AMCOM via e-mail, fax, or the World Wide Web. Our fax
number is DSN 788-6546 or Commercial 256-842-6546. Our e-mail address is
[email protected]. Instructions for sending an electronic 2028 may be found
at the back of this manual. For the World Wide Web, use
https://amcom2028.redstone.army.mil.
Paragraph Page
SECTION I. IDENTIFICATION AND DESCRIPTION
Test instrument identification .............................. 1 2
Forms, records, and reports .................................. 2 2
Calibration description .......................................... 3 3
II. EQUIPMENT REQUIREMENTS
Equipment required............................................... 4 4
Accessories required .............................................. 5 4
III. CALIBRATION PROCESS FOR TEKTRONIX,
TYPE SC501
Preliminary instructions ....................................... 6 5
Equipment setup.................................................... 7 5
Input compensation ............................................... 8 6
Vertical gain ........................................................... 9 7
Timing..................................................................... 10 8
Bandwidth .............................................................. 11 9
Power supply .......................................................... 12 9
Final procedure ...................................................... 13 9
___________
*This bulletin supersedes TB 9-6625-2173-35, dated 24 March 2003.
TB 9-6625-2173-35
Paragraph Page
SECTION IV. CALIBRATION PROCESS FOR TEKTRONIX,
TYPE SC502
Preliminary instructions........................................ 14 10
Equipment setup .................................................... 15 10
Preamplifier input capacitance ............................. 16 11
Volts/div accuracy................................................... 17 11
Horizontal centering and sweep gain ................... 18 13
X10 magnifier gain and fast timing ...................... 19 13
Vertical output amplifier ....................................... 20 15
Vertical bandwidth................................................. 21 15
Calibrator................................................................ 22 16
Power supply .......................................................... 23 16
Final procedure ...................................................... 24 17
V. CALIBRATION PROCESS FOR TEKTRONIX,
TYPE SC504
Preliminary instructions........................................ 25 17
Equipment setup .................................................... 26 18
Calibrator................................................................ 27 18
Vertical gain ........................................................... 28 20
Horizontal X gain ................................................... 29 21
Vertical amplifier transient response ................... 30 22
Vertical bandwidth................................................. 31 23
External triggering sensitivity .............................. 32 23
Sweep timing .......................................................... 33 25
Power supply .......................................................... 34 28
Final procedure ...................................................... 35 28
SECTION I
IDENTIFICATION AND DESCRIPTION
1. Test Instrument Identification. This bulletin provides instructions for the
calibration of Oscilloscope, Tektronix, Types SC501, SC502, and SC504. The
manufacturers' manuals were used as the prime data sources in compiling these
instructions. The equipment being calibrated will be referred to as the TI (test instrument)
throughout this bulletin.
a. Model Variations. Variations among models are indicated in the text.
b. Time and Technique. The time required for this calibration is approximately 1
hour, using the dc and low frequency technique.
2. Forms, Records, and Reports
a. Forms, records, and reports required for calibration personnel at all levels are
prescribed by TB 750-25.
2
TB 9-6625-2173-35
b. Adjustments to be reported are designated (R) at the end of the sentence in which
they appear. When adjustments are in tables, the (R) follows the designated adjustment.
Report only those adjustments made and designated with (R).
3. Calibration Description. TI parameters and performance specifications which
pertain to this calibration are listed in table 1.
Table 1. Calibration Description
Test instrument parameters Performance specifications
Type SC501
Vertical sensitivity Range: 10,100 and 1000 mV/div
Accuracy: